Page 64 - International safety guide for oil tankers and terminals
P. 64

INTERNATIONAL SAFETY GUIDE FOR OIL TANKERS AND TERMINALS




                                      Releasing nitrogen or carbon dioxide in a confined or unventilated
      ---ใช้เพื่อการศึกษาเท่านั้น---
                                      area can lower the concentration of oxygen to a level that is
                                      immediately dangerous to life or health. Calibration should
                                      therefore only be carried out in well ventilated areas.
               งานห้องสมุด ศูนย์ฝกพาณิชย์นาวี

                                    The analyser readings are directly proportional to the pressure in the
                                    measuring cell. The unit is calibrated to a specific atmospheric pressure
                                    and the small error due to atmospheric pressure variations can be
                                    corrected if required. Continuous samples should be supplied to the
                                    instrument by positive pressure. They should not be drawn through the
                                    analyser by negative pressure as the measuring pressure then becomes
                                    uncertain.

                                    The filter should be cleared or replaced when an increase in sample
                                    pressure is required to maintain a reasonable gas flow through the
                                    analyser. The same effect is produced if the filter becomes wet due to
                                    insufficient gas drying. The need for filter cleaning or replacement should
                                    be checked regularly.


                         2.4.10.2 Electrochemical Sensors
                                    Analysers of this type determine the oxygen content of a gas mixture by
                                    measuring the output of an electrochemical cell. In one commonly used
                                                                          ึ
                                    analyser, oxygen diffuses through a membrane into the cell, causing
                                    current to flow between two special electrodes separated by a liquid or
                                    gel electrolyte.

                                    The current flow is related to the oxygen concentration in the sample and
                                    the scale is arranged to give a direct indication of oxygen content. The
                                    cell may be housed in a separate sensor head connected by cable to the
                                    read out unit.

                                    The analyser readings are directly proportional to the pressure in the
                                    measuring cell, but only small errors are caused by normal variations in
                                    atmospheric pressure.

                                    Certain gases may affect the sensor and give rise to false readings.
                                    Sulphur dioxide and oxides of nitrogen interfere if they are present in
                                    concentrations of more than 0.25% by volume. Mercaptans and hydrogen
                                    sulphide can poison the sensor if their levels are greater than 1% by
                                    volume. This poisoning does not occur immediately but over a period of
                                    time; a poisoned sensor drifts and cannot be calibrated in air. In such
                                    cases, reference should be made to the manufacturer’s instructions.

                         2.4.10.3 Maintenance, Calibration and Test Procedures

                                    As these oxygen analysers are of vital importance, they should have a
                                    valid calibration certificate and should be tested strictly in accordance
                                    with the manufacturer’s instructions before use.

                                    It is essential that, each time an instrument is to be used, a check is
                                    made of batteries (if fitted) and zero point (21% oxygen) setting. During
                                    use, frequent checks should be made to ensure accurate readings are
                                    obtained at all times.

                                    Testing is simple on all analysers using atmospheric air to test the
                                    reference point (21% oxygen) and an inert gas to test the 0% oxygen




       30                                      © ICS/OCIMF/IAPH 2006
   59   60   61   62   63   64   65   66   67   68   69